Easychem Coli Online - Coliform & E coli analyser - Systea Datasheet
EASYCHEM COLI ONLINE
MEASURING PRINCIPLE: Colorimetric (Total & Faecal Coliforms) and fluorimetric (E. coli)
MEASUREMENT TYPE: Fully automated, direct read, on-line discrete analyser for Presence/Absence (P/A) and Most Probable Number (MPN)
MEASUREMENT SENSITIVITY:
1 CFU / 100 mL (P/A)
1.8 – 1,250 up to 17,500 MPN /100 mL (MTF)
Other measurement ranges are programmable, according to the expected concentrations and the analytical frequency
MPN MEASUREMENT
FREQUENCY Up to 12 samples / day (Range depending on customer set-up)
MPN MEASUREMENT
TIME: 15-18 hours
MEASURING POINTS: One, multipoint as option
OPERATOR INTERFACE: Colour graphic touch-screen 15” display
Thiết bị đo COD, BOD, TOC
Xem tất cả
Easychem Coli Online - Coliform & E coli analyser - Systea Datasheet
Free preview
EASYCHEM COLI ONLINE
MEASURING PRINCIPLE: Colorimetric (Total & Faecal Coliforms) and fluorimetric (E. coli)
MEASUREMENT TYPE: Fully automated, direct read, on-line discrete analyser for Presence/Absence (P/A) and Most Probable Number (MPN)
MEASUREMENT SENSITIVITY:
1 CFU / 100 mL (P/A)
1.8 – 1,250 up to 17,500 MPN /100 mL (MTF)
Other measurement ranges are programmable, according to the expected concentrations and the analytical frequency
MPN MEASUREMENT
FREQUENCY Up to 12 samples / day (Range depending on customer set-up)
MPN MEASUREMENT
TIME: 15-18 hours
MEASURING POINTS: One, multipoint as option
OPERATOR INTERFACE: Colour graphic touch-screen 15” display
Thiết bị đo Phosphate, Total Nitrogen, Total Phosphorus
Xem tất cả
MicC_uLFR HT DR - TP - Micromac Total Phosphorus - Systea Datasheet
Free preview
MICROMAC PHOSPHORUS TOTAL is a microprocessor controlled On Line analyzer specifically designed for automatic Total Phosphorous monitoring on several types of water matrix.
ROBUST AND RELIABLE
Designed for industrial and Environmental On Line applications ensures the highest level of robustness in the electronics, mechanics and hydraulics components.
Complete separation between electronics and hydraulics plus a simple and robust LFA * hydraulics allows easy maintenance and long terms reliable operations.
* LFA: Loop Flow Analysis patent pending
MicC_uLFR HT FC - TN@220nm - Micromac Total Nitrogen - Systea Datasheet
Free preview
MICROMAC TOTAL NITROGEN is a microprocessor controlled On Line analyzer specifically designed for automatic Total Nitrogen monitoring on several types of water matrix.
ROBUST AND RELIABLE
Designed for industrial and Environmental On Line applications Micromac C ensures the highest level of robustness in the electronics, mechanics and hydraulics components. Complete separation between electronics and hydraulics plus a simple and robust LFA * hydraulics allows easy maintenance and long terms reliable operations.
*µLFR: µ Loop Flow Reactor - Patent pending
MicC_Phosphate - Micromac Phosphate - Systea Datasheet
Free preview
MICROMAC PHOSPHATE is a microprocessor controlled On Line analyzer specifically designed for automatic phosphate monitoring on several types of waters matrix.
ROBUST AND RELIABLE
Designed for industrial and Environmental On Line applications ensures the highest level of robustness in the electronics, mechanics and hydraulics components.
Complete separation between electronics and hydraulics plus a simple and robust LFA * hydraulics allows easy maintenance and long terms reliable operations.
* LFA: Loop Flow Analysis patent pending
Máy phân tích kim loại nặng trong nước
Xem tất cả
MicCfa_Cyanide Total - Micromac Cyanide Total - Systea Datasheet
Free preview
MICROMAC CYANIDE TOTAL is a microprocessor controlled On Line analyzer specifically designed for automatic Cyanide monitoring on several types of water matrices.
ROBUST AND RELIABLE
Designed for industrial and Environmental On Line applications ensures the highest level of robustness in the electronics, mechanics and hydraulics components.
Complete separation between electronics and hydraulics plus a simple and robust batch type CFA* hydraulics allows easy maintenance and long terms reliable operations.
*CFA: Continuous Flow Analysis
MicC_uLFR HT DR - TFe - Micromac Total Iron - Systea Datasheet
Free preview
MICROMAC TOTAL IRON is a microprocessor controlled On Line analyzer specifically designed for automatic Total Iron monitoring on several types of
waters matrix.
ROBUST AND RELIABLE
Designed for industrial and Environmental On Line applications Micromac C ensures the highest level of robustness in the electronics, mechanics and hydraulics components. Complete separation between electronics and hydraulics plus a simple and robust LFA* hydraulics allows easy maintenance and long terms reliable operations.
*LFA: Loop Flow Analysis patent pending
MicC_Chromium6+ - Micromac Chromium 6+ - Systea Datasheet
Free preview
MICROMAC CHROMIUM HEX is a microprocessor controlled On Line analyzer specifically designed for automatic Chromium 6+ monitoring on several types of water matrices.
ROBUST AND RELIABLE
Designed for industrial and Environmental On Line applications ensures the highest level of robustness in the electronics, mechanics and hydraulics components.
Complete separation between electronics and hydraulics plus a simple and robust LFA* hydraulics allows easy maintenance and long terms reliable operations.
*LFA: Loop Flow Analysis patent pending
MicC_uLFR HT - TMn - Micromac Manganese Total - Systea Datasheet
Free preview
MICROMAC TOTAL MANGANESE is a microprocessor controlled On Line analyzer specifically designed for automatic Total Manganese monitoring on several types of water matrix.
ROBUST AND RELIABLE
Designed for industrial and Environmental On Line applications Micromac C ensures the highest level of robustness in the electronics, mechanics and hydraulics components. Complete separation between electronics and hydraulics plus a simple and robust LFA* hydraulics allows easy maintenance and long terms reliable operations.

* LFA: Loop Flow Analysis patent pending

Technical Specifications
MEASURING PRINCIPLE: Colorimetric, Formaldoxine method, combined UV + HT digestion
COLORIMETER: dual beam, silicon detector
MEASUREMENT TYPE: cyclic; MEASURING INTERVAL: programmable
MEASURING TIME: 13-15 minutes depending on the ragem
MEASURING RANGE: 0-2 ppm Mn, other ranges available on request
DETECTION LIMIT: typically better 3% of the full scale, calculated as for EPA p. 136 appendix B
REPEATABILITY: better than 5%; SD% of 7 replicates at 50% of the full scale
ACCURACY: better than 5% 7 replicates at 20% of the full scale
OUTPUT SIGNAL: 4-20 mA INPUT SIGNALS: n. 1 Analysis, n. 1 calibration; digital contacts
ALARMS: n. 1 High Limit, n. 1 General, n. 1 Calibration; potential free contacts
SAMPLE AND WASTE DELIVERY: pressure free; SAMPLE TEMPERATURE: 10 °C- 30 °C
REAGENTS REPLACEMENT: 3/4 weeks depending on the operating temperature
PROTECTION: IP55
HARDWARE: PC104 industrial standard, Integrated keyboard and graphics display, RS232 option
POWER SUPPLY: 12 V DC external power supply included; 4W Standby; 50 W (mean) analysis
WEIGHT: 33 Kg without reagents; DIMENSION: 800x420x280 mm(hxwxd)